3 rear panel, Warning - risk of electric shock – INFICON SQC-310 Thin Film Deposition Controller User Manual

Page 15

Advertising
background image

1 - 3

IP

N 07

4-

55

0-

P1

B

SQC-310 Operating Manual

1.3 Rear Panel

Figure 1-3 Rear Panel

Table 1-2 Rear Panel Connections

Sensor 1 & 2

Connects to the oscillator. See

section 1.4 on page 1-4

.

Output 1 & 2

Connects the SQC-310 output to your evaporation supply control
input (see next section).

I/O (1-8)

Connects 8 relays and 8 digital inputs to external equipment for
process control. See

Appendix C

.

RS-232
USB or Ethernet

Connects to a computer for programming and data acquisition.
RS-232 and USB are standard. Ethernet option replaces USB.

Sensor 3 & 4,
Output 3 & 4,
I/O 9-16

Increases the number of input, output, and digital I/O connections
when the optional expansion card is installed.

Measurement ground terminal useful for common system and
cable grounding.

Power Input and
Fuse

Connects to mains power. The SQC-310 automatically detects
main voltages of 100-120 and 200-240 V (ac), 50/60 Hz

WARNING - Risk Of Electric Shock

For continued protection, replace fuses
with the proper type and rating. Use
power cords only of the specified type
and rating, attached to a properly
grounded receptacle.

100-120/200-240 V~

50/60 Hz

25 VA

Sensor 1

Sensor 2

Output 1

Output 2

I/O 1-8

RS-232

Sensor 3

Sensor 4

Output 3

Output 4

I/O 9-16

Fuse T.5A 250V

SQC-310 Deposition Controller

Serial No.

USB/Ethernet