9 get/set sensor, source, relay parameters – INFICON SQC-310 Thin Film Deposition Controller User Manual

Page 91

Advertising
background image

5 - 11

IP

N 07

4-

55

0-

P1

B

SQC-310 Operating Manual

Parameters:

1

Turn Relay: On (locks relay for remote use only)

2

Turn Relay: Off (locks relay for remote use only)

3

Unlock Relay (Releases the Relay back to the Instrument)

Format: . . . . . . . . GE[Relay#] [Parameter1]

Example: . . . . . . GE1 1

Returns: A

5.3.9 Get/Set Sensor, Source, Relay Parameters

Command: . . . . . H (obsolete)

Description: . . . . Gets/Sets the logical function of each of the 16 relays.

Parameters:

1

Source 1 Shutter

21

Max. Power

2

Source 2 Shutter

22

Stopped

3

Source 3 Shutter

23

Time Setpoint

4

Source 4 Shutter

24

Thick Limit

5-8

Sensor 1-4 Shutter

25

Final Thickness

9-12

Sensor 1-4 Fail

26

Dual Xtal 1/2 Shutter

13

All Crystals Good

27

Dual Xtal 3/4 Shutter

14

All Crystals Bad

28-35

Out1 Pocket 1-8

15

Process Hold

36-43

Out1 Pocket 1-8

16

Deposit Phase

44-51

Out1 Pocket 1-8

17

PreCond Phase

52-59

Out1 Pocket 1-8

18

SoakHold Phase

19

Process Active

20

Manual Mode

Example: . . . . . . To Get Relay 1 and Relay 2 functions, send: H? 1 2

The return for Source 1 Shutter and Sensor 1 Shutter would be:
1,1 2,5
To Set the same parameters you would send: H1 1,1 2,5

Command: . . . . . HA

Description: . . . . Gets/Sets the parameters of each of the 2 or 4 sensors.

Parameters:

1

Crystal Tooling

6

Drive Type

2

Number of Positions

7

Feedback Type

3

Shutter

8

Indexer Delay (sec.)

4

Shutter Delay (ms)

9

Current Crystal Position

5

Control Type