3.11.3 Sensors and Sources Menu

The Sensors and Sources Menu allows you to configure the SQC-310 to the types
of sensors and sources in your system, particularly multi-crystal sensors and
multi-pocket sources.

Figure 3-20 Sensor 1 Edit

3.11.3.1 Sensor Setup

Crystal Tooling: Adjusts for the difference in measured deposition rate between
the sensor and the substrate being coated.

Figure 3-21 Tooling

In the left illustration, the sensor will measure less rate or thickness than is actually
deposited on the substrate because of its positioning. In the right illustration, the
sensor will measure more. Tooling is the ratio of the actual substrate deposition
rate or thickness, to that measured by the sensor. If the rate/thickness reading is
low, then increase the tooling value. If the rate/thickness reading is high, then lower
the tooling value.

Substrate

Substrate

Tooling

Over 100%

Tooling

Under 100%