2.6 Sensor and Source Setup

The SQC-310 must be configured to match the type of sensor installed in your
system. We will set up a single sensor without a shutter.

Chapter 3

discusses the

other Sensor configuration options.

Select System Menu, Sensors & Sources, scroll to Sensor 1 and press Select to
Display the Sensor 1 menu. See

Figure 2-8

.

Figure 2-8 Sensor 1 Edit

Crystal Tooling adjusts for the difference in measured deposition rate between the
sensor and the substrate being coated.

Figure 2-9 Tooling Over/Under

In the left illustration in

Figure 2-9

, the sensor will measure less rate or thickness

than is actually deposited on the substrate because of its positioning. In the right
illustration, the sensor will measure high. Tooling is the ratio of the actual substrate
deposition rate or thickness, to that measured by the sensor.

A simple rule to remember is: If the rate/thickness reading is low, then increase the
tooling value. If the rate/thickness reading is high, then lower the tooling value.

Crystal Tooling

Number of Positions

2. Sensor 2
1. Source 1

2. Source 2

Indexer

Single Source

Single Crystal

1

100%

To

Main

Prev

Menu

Done

Sensors & Sources Menu

Name

Value

1. Sensor 1

Shutter

No

Edit

Substrate

Substrate

Tooling

Over 100%

Tooling

Under 100%