SQC-310 Operating Manual

5.3.7 Get/Set Material Parameters

Command: . . . . . F

Description: . . . . Gets/Sets the parameters of the 100 stored materials.

Parameters:

1.

Material Name

3.

Zfactor

2.

Density

Example: . . . . . . To Get Material 1 name, send: F1? 1

The return string would be of the form: Aluminum
To Set the parameter you would send: F1 1,AnyName

NOTE: You can only get/set one material parameter at a time.

5.3.8 Get/Set Input & Relay Parameters

Command: . . . . . G
Description: . . . . Gets/Sets the logical function of each of the 16 digital inputs.
Parameters:

1

Start Process

32

Soak Hold

2

Abort Process

33

Zero Thickness

3

Start Layer

34

Zero Time

4

Stop Layer

35

Out1 Pocket Ready

5

Start Next Layer

36

Out2 Pocket Ready

6

Force Final Thick

37

Out3 Pocket Ready

7-31

Start Process 1-25

38

Out4 Pocket Ready

39

Not Used

Example: . . . . . . To Get Input 1 and Input 2 functions, send: G? 1 2

The return string for Start Layer and Stop Layer would be:
1,3 2,4

To Set the same parameters you would send: G1 1,3 2,4

Command: . . . . . GA
Description: . . . . Gets/Sets the parameters of each of the 8 or 16 digital inputs.
Parameters:

1

Name

3

Pin Number

2

Active Level

4

Input In Use *

5

NameMode

Format: . . . . . . . . GA[Input#]? [Parameter1] [Parameter2] …
Example: . . . . . . GA1? 2 3

Returns: !A2,0 3,1
NOTE: When changing the name (GA1), you MUST also send

the Name Mode command (GA5) as the next
parameter.