Metrohm tiamo 1.0 Manual User Manual

Page 136

Advertising
background image

112

Workplace

4 - Preconditioning (only for methods with conditioning)

In methods that contain commands with switched-on conditioning and in which

the option

Automatic conditioning

is switched on in the

START

command then pre-

conditioning will start for all these commands. Preconditioning does not belong to
the determination itself and does not increase any counter (

Sample number

,

Statis-

tics

).

• Conditioning requirement not fulfilled

After the start of preconditioning the status changes to

COND BUSY

. In

this status conditioning can be started with

[Start]

even though the

conditioning requirement has not been fulfilled. Conditioning can also
be canceled with

[Stop]

or interrupted with

[Hold]

and then continued

with

[Cont]

. If an error occurs during conditioning then neither the Error

track nor the Exit track will be carried out.

• Conditioning requirement fulfilled

If the conditioning requirement has been fulfilled then the status
changes to

COND READY

. In this status the determination can be

started with

[Start]

. Conditioning can also be canceled with

[Stop]

or

interrupted with

[Hold]

and then continued with

[Cont]

. The condition-

ing requirement will continue to be monitored continuously until the
determination has been started.

5 - Start method

At the start of the method the following checks and actions are triggered:

• Check sample data

Checks that the entered Sample data corresponds to that defined in
the Method variables and is valid.

• Assign determination ID

An unambiguous determination ID is assigned for the determination.

• Reserve device (module)

The devices (or their functional units) used by the device-dependent

commands contained in the method are reserved for the duration of
the determination, i.e. they cannot be used in other workplaces nor
can they be operated manually.

• Increment sample number

The

Sample number

is increased by

+1

.

• Increment statistics counter

If statistics has been switched on in both the

START

command and on

the tab

Single determinations

then the statistics counter is increased by

+1

. If a new method is loaded or if the statistics counter agrees with

the set counter then the current statistical data will first be deleted and

the statistics counter reset to

0

.

6 - Main run

When processing the loaded method the following actions can be carried out:

• Process main track

After the start of the Main track the status changes to

BUSY

. The com-

mands of the main track and the other tracks that it calls up are proc-

essed in sequence. If the main track calls up tracks that are not free
then a pause will be made until these tracks can be started.

• Interrupt and continue run

A running determination can be interrupted at any time with

[Hold]

and

then continued with

[Cont]

. All the active tracks will be stopped and

continued together.

Advertising