Continuous operation – Metrohm tiamo 1.0 Manual User Manual
Page 92

68
General program functions
Continuous operation
Stirrers can be switched on for a defined period here. This tab appears for the se-
lection of a single stirrer as well as for all stirrers.
Stirring rate
-15 ... -1, 1 ... [ 8 ] ... 15
Selection of the stirring speed. This parameter can also be altered live.
Stirring period
1 ... [ 60 ] ... 999999 s
Entry of the time during which stirring is to take place. If this parameter is al-
tered after stirring has started then it will only apply to the next stirring pro-
cedure.
Start continuous operation for the selected stirrer(s). The remaining time is
shown in the status display. When the stirring period has expired the stirrer
will be switched off automatically.
Stop continuous operation for the selected stirrer(s).