Metrohm tiamo 1.0 Manual User Manual

Page 342

Advertising
background image

318

Method

Start conditions

Command name

25 characters, [ DET pH # ]

Name of the command.

The start conditions are processed in the listed sequence before the titration is
started.

Initial measured value

The initial measured value is determined before the other start conditions are

processed. The measured value acceptance depends on the three following pa-
rameters:

Signal drift

0.1 ... 999.0 mV/min, [ off ]

(only Titrando)

The measured value is only accepted if the drift undercuts the entered
value.

off

means that the measured value is accepted after the maximum

waiting time.

Min. waiting time

[ 0 ] ... 999999 s

(only Titrando)

The minimum waiting time is only important for drift-controlled measure-
ments. The measured value is accepted only after the minimum waiting

time has elapsed, even if the signal drift has already been achieved. The
drift continues to be checked as the minimum waiting time is passing.

Max. waiting time

0 ... [ 1 ] ... 999999 s

(only Titrando)

If signal drift has been switched off or has not yet been achieved then the
measured value will be accepted when the maximum waiting time has

elapsed.