Metrohm tiamo 1.0 Manual User Manual
Page 395

Commands
371
Max. waiting time
0 ... [ 1 ] ... 999999 s
(only Titrando)
If signal drift has been switched off or has not yet been achieved then the
measured value will be accepted when the maximum waiting time has
elapsed.
Pause 1
Pause 1
[ 0 ] ... 999999 s
(only Titrando and Titrino 719, 720, 736, 751, 758, 785, 794,
798, 799)
Waiting time, e.g. until the electrode has settled down, before a start vol-
ume is added.
Start volume
Start volume
[ 0.00000 ] ... 9999.99 mL
(Titrando),
[ 0.00 ] ... 999.99 mL
(Titrino)
Volume to be added before the start of the titration with the below entered
speed.
Dosing rate
0.01 ... 166.00 mL/min, [ maximum ]
(Titrando),
0.01 ... 150.00 mL/min, [ maximum ]
(Titrino)
Speed at which the start volume is to be added. The maximum dosing rate
depends on the cylinder volume of the used exchange unit or dosing unit.
Pause 2
Pause 2
[ 0 ] ... 999999 s
Waiting time, e.g. until the electrode has settled down after the start or a re-
action time after the addition of a start volume. The pause takes place at
the end of all start conditions.