Commands

375

Command name

25 characters, [ SET pH # ]

Name of the command.

Conditions for stopping the titration. The stop conditions are checked at the same
time and the condition which is fulfilled first stops the titration.

Stop volume

0.00000 ... [ 100.000 ] ... 9999.99 mL, off

(Titrando)

0.00 ... [ 100.00 ] ... 9999.99 mL, off

(Titrino)

Stops when the given volume (including start conditions) has been added
after the start of the titration. The stop volume should be adapted to suit the
sample weight or the titration vessel size.

Stop time

0 ... 999999 s, [ off ]

(only Titrando)

Stops when the preset time (including start conditions) has elapsed since
the start of the titration.

Filling rate

0.01 ... 166.00 mL/min, [ maximum ]

(Titrando)

0.01 ... 150.00 mL/min, [ maximum ]

(Titrino)

Speed with which the buret is to be refilled after the titration. The maximum
filling rate depends on the cylinder volume of the exchange unit or dosing
unit used.

Conditioning

Command name

25 characters, [ SET pH # ]

Name of the command.