Error track – Metrohm tiamo 1.0 Manual User Manual
Page 324

300
Method
manually. If a new exit trackis created a
DATABASE
and a
REPORT
command are
automatically inserted in this track. With these commands it can be guaranteed
that a data base entry and a report are created for each determination even if the
determination has been stopped either manually or due to an error.
Error track
An error track is a track that can be created manually with
Insert, New track, Error
track
or with the icon . It starts with a start command
ERROR
and finishes with a
end command
END
. A method can only contain one error track.
The error track is executed if the determination is stopped due to an error. In this
case the error track takes over the function of the Main track. Normal tracks can
be called from error tracks as well.