Metrohm tiamo 1.0 Manual User Manual
Page 333
Commands
309
Method variables (properties)
Definition of the method variable
Name
50 characters
Name of the method variable which must be unambiguous inside the
method.
Type
[ Text ], Number, Date/Time
Selection of the type of the variable. This parameter is inactive for the
method variables
Sample Size
,
Sample size unit
,
Sample position
.
Assignment
[ on ], off
ID1 ... ID8, Sample size, Sample size unit, Sample position
If this option is enabled, a sample variable can be selected which will be
assigned to this method variable. Only those variables are selectable which
are not yet assigned. Sample data variables not assigned to a method vari-
able are not selectable in other commands and are not stored in the deter-
mination either. No formulae can be entered in this field.
Note
For the sample data variables
ID1 ... ID8
,
Sample size
,
Sample size unit
,
Sample
position
the name of the assigned method variable will be shown automatically
on the tab
Single determination
in the window
Run
as well as on the tab
Sample
in
the subwindow
Information
.