Metrohm tiamo 1.0 Manual User Manual

Page 414

Advertising
background image

390

Method

Command name

25 characters, [ SET Upol # ]

Name of the command.

The following additional methods for evaluation of titration curves can be activated
and defined here:

Fix end point evaluation

on, [ off ]

For a fixed quantity (measured value, volume or time) for the fix end point

the associated values are interpolated from the list of measuring points.

Fixed quantity

[ Measured value ], Time, Volume

Selection of the fixed quantity to which the associated values for the other
quantities are to be interpolated from the list of measured points.

Fix EP# at

-200.0 ... 200.0 µA, [ off ]

Fix end point 1 ... 9 for

Fixed quantity

=

Measured value

.

Fix EP# at

0.0 ... 999999.9 s, [ off ]

Fix end point 1 ... 9 for

Fixed quantity

=

Time

.

Fix EP# at

0.00000 ... 9999.99 mL, [ off ]

Fix end point 1 ... 9 for

Fixed quantity

=

Volume

.

With this button additionally to the always shown fix EP1...3 the fix EP4...9
are displayed.

With this button instead of fix EP1...9 only the fix EP1...3 are shown.

Minimum evaluation

on, [ off ]

For the minimum measured value the associated volume, time and tem-
perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation
).

Threshold value

0.5 ... [ 5.0 ] ... 10.0 µA/mL

The evaluation begins as soon as the slope of the curve exceeds the set
threshold value.

Maximum evaluation

on, [ off ]

For the maximum measured value the associated volume, time and tem-

perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation
.

Threshold value

0.5 ... [ 5.0 ] ... 10.0 µA/mL

The evaluation begins as soon as the slope of the curve exceeds the set
threshold value.

Additional measured values

Screenshot and parameters: see DET pH - Additional measured values