Determination run – Metrohm tiamo 1.0 Manual User Manual

Page 145

Advertising
background image

Subwindow Run

121

Dark orange background

Selected currently running line. If sample data is only altered after the start

of the determination (in the status

BUSY

) then this is regarded as being a

live modification.

White background

Lines that have not yet been processed. These lines can be edited.

Turquoise background

Selected line that has not yet been processed.

Loaded

Shows the sample table whose data has been loaded into the working
sample table with

Sample table, Load...

. If no sample table has been loaded

or if all the lines have been deleted then this field will be empty. If data in a
loaded sample table is later altered or if new lines are added then in addi-
tion to the name

(modified)

will be shown.

For possible ways of editing the working sample table and the window menu
please refer to Edit working sample table.

Determination run

In a series of determinations that are started on the tab

Determination series

in the

subwindow

Run

each determination takes place as follows:

1 - Load sample data

The sample data entered in the working sample table for the current line will be

loaded for the determination.

2 - Select method

Method selection triggers the following actions:

Select method without assignment ID

If the option

Method selection by assignment ID

is not switched on the

tab

Process

then the method selected in the field

Method

will remain

loaded.

Select method with assignment ID

If the option

Method selection by assignment ID

is switched on the tab

Process

then the

Sample assignment ID

will be used to load the method

defined in the

Sample assignment table

.

3 - Carry out start test

During the Start test the following checks and actions are triggered:

Right to carry out the method?

Checks whether the logged-in user has the right to execute the par-
ticular method.

Method executable?

Checks whether the particular method is executable. If the method is

non-executable then a message appears asking for the method test to
be carried out and the method edited accordingly.

Check device (module) and rack

Checks whether the devices (or device types), modules (e.g. measur-
ing inputs, towers, etc.) and racks demanded in the device-specific

commands are present and available or assigned. If the device has
not yet been assigned in a command then the window

Device assign-

ment

appears for assigning the device.