Metrohm tiamo 1.0 Manual User Manual

Page 436

Advertising
background image

412

Method

[ automatic ]

If a temperature sensor is connected, the temperature will be measured

continuously. Otherwise the temperature manually entered under

Tempera-

ture

(see Measuring parameters) will be used.

off

The temperature will not be measured. The temperature manually entered
under

Temperature

(see Measuring parameters) will be used.

Stirrer

Stirrer

[ 1 ] ... 4, off

(Titrando only)

Selection of the stirrer.

off

means that no stirrer will be used.

Stirring rate

-15 ... [ 8 ] ... 15

(Titrando only)

Setting the stirrer speed. The direction in which stirring takes place changes
as the sign in front of the stirrer speed changes.

Switch off automatically

[ on ], off

(Titrando only)

The stirrer is automatically switched off at the end of the command if this
option is enabled.

Measuring parameters

Command name

25 characters, [ MEAS pH # ]

Name of the command.

The measuring parameters are defined on this tab.

Measurement

The two measuring modes Measurement with drift control or Meas-
urement without drift control
are selectable.