Metrohm tiamo 1.0 Manual User Manual
Page 462

438
Method
Evaluations
Command name
25 characters, [ MEAS T/Flow # ]
Name of the command.
The following methods for evaluating measuring curves are defined on this tab.
Minimum evaluation Temperature
on, [ off ]
For the minimum measured value the associated time and temperature are
interpolated from the list of measuring points) (see Titration commands -
Evaluation - Minimum/Maximum evaluation).
Threshold value
0.1 ... [ 1.0 ] ... 20.0 °C/s
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the threshold value.
Maximum evaluation Temperature
on, [ off ]
For the maximum measured value the associated time and temperature are
interpolated from the list of measuring points) (see Titration commands -
Evaluation - Minimum/Maximum evaluation).
Threshold value
0.1 ... [ 1.0 ] ... 20.0 °C/s
The evaluation of the maximum begins as soon as the slope of the curve
exceeds the threshold value.
Additional measured values
Screenshot and parameters: see DET pH - Additional measured values