Subwindow Devices
615
Temperature correction
Shows the temperature correction set on the device. This parameter directly
affects the temperature control and allows it to be adjusted. In this way any
temperature difference that may occur between the oven and sample can
be compensated. The displayed temperature is the oven temperature cor-
rected by this value.
Gas
On the tab
Gas
the parameters for the gas flow to the 774 Oven Sample Proces-
sor are shown. These parameters cannot be edited; they are only read off by the
device.
Unit for gas flow
Shows the unit selected at the device for the gas flow display (
mL/min
or
L/h
).
Minimum gas flow
Shows the lower warning limit set on the device for the gas flow.
Maximum gas flow
Shows the upper warning limit set on the device for the gas flow.
Gas type
Shows the carrier gas.
Correction factor
Shows the correction for the gas flow measurement with "other" gases.
Note
If the minimum gas flow is undercut or the maximum gas flow is exceeded then
an error message will appear.