Immunohistochemical Report (IHC Report)

1. From

the

PROGRAMMING GRID, select the Print button. The WHICH REPORT? screen

appears.

2. Select

the

IHC button.

The IHC Report can also be printed from the RUN LOG screen after the completion of a
run.

3. After a run is completed, a PRINT IHC button appears at the bottom of the screen.

4. Select the PRINT IHC button.

The IHC Report format is selected during Autostainer initialization. This format can be changed
when necessary by selecting the OPTIONS button from the INITIALIZE screen. (see Initializing the
Dako Autostainer, Section 6)

The IHC Report can be configured to print in one of the following formats:

• The One case/page option prints one case on each page of the report.

• The One slide ID/page option prints each Slide ID on a separate page.

• The No page break option prints all slides for a run in a continuous print out.

The IHC Report groups all slides by case number and prints the following information for every
slide:

• slide

#

• primary antibody information (Full name, Species reactivity, Catalog #, Lot # (if entered),

Expiration date, and Incubation time)

• full name and incubation time for all reagents used

Run Log Printout

During a staining run a Run Log is created containing each step performed during the run. The
run log can be printed at the end of the run after saving the Run Log (see Completing a Run,
section 14). The Run Log can also be printed from the Programming Grid:

1. From

the

PROGRAMMING GRID, select the Print button. The WHICH REPORT? screen

appears.

2. Select

the

RUN LOG button.