Section 13

|

Loading Slides

After programming is completed and the reagents have been loaded, you are ready to load the slides
on the slide racks. Print the Slide Layout Map by selecting the PRINT button in the SLIDE LAYOUT
MAP:LOAD SLIDES
screen. This map provides a guide for loading slides and can be saved as part
of your staining record.

Loading Microscope Slides

NOTE: The slides should soak in a buffer solution containing 0.05% Tween 20 for a minimum of 5
minutes prior to loading on the Autostainer.

1. Tilt the slide rack at a 30

°

angle away from you, allowing the back of the rack handles to

rest in the loading slots.

2. Hold the slide by placing your thumb and forefinger on each side of the slide’s unfrosted

end.

3. Load the slide into the slide rack by gently pushing the frosted end into position. The slide

is securely held to the rack. (see photo)

4. Secure the slide rack in the instrument by aligning the holes in the handles over the pins

in the sink frame. Gently flood the slides with buffer containing 0.05% Tween 20 using a
squirt bottle to prevent the slides from drying while loading additional slides.





















Microscope slide

Slide Rack

Frosted end
of slide