Troubleshooting

Symptom

Possible Causes

Solutions

Autostainer Plus will not turn
on (top green LED not on). A
Cannot Continue box displays
“Instrument Not On” when a
run is started.

The Autostainer Plus is
unplugged at the power source.

Check the power cord
connections and verify the
surge protector is plugged in
and turned on.

The Autostainer Plus does not
respond when you initiate a
prime pump or a run. (bottom
LED not on) A Cannot
Continue box displays
“Instrument Not On” when a
run is started.

The cable between the
Autostainer Plus and the
computer is not connected
properly.

Shut down the system and turn
the power off. Disconnect and
reconnect the cable between
the Autostainer Plus and the
computer. ** Please take great
care in this action. Pins can be
bent and cause damage to the
instrument during operation.

Autostainer Plus suddenly
stops during a run.

The Autostainer Plus will stop
frequently during the run. This
is due to set incubation times.
Check the run log to verify.

No action is needed. The
Autostainer Plus will resume
movement. Check Run Log
screen.