Paused to add reagent, Slides skipped, Partial reagent application – Dako Autostainer Plus User Manual

Page 79

Advertising
background image

Dako Autostainer Plus

|

User Guide

79

reagent vial. If sufficient volume of a reagent is not available, the Autostainer Plus will provide a list on
the Run Log of skipped or partially applied slides.

To program the Autostainer Plus to check for insufficient volumes at the start of the run, select the
Check volumes option on the SET START TIME screen (see Starting a Run, section 14).

Paused to Add Reagent

Reagent vials are loaded on the Autostainer Plus and a run has been started.

1. The reagent probe senses insufficient or a missing reagent. The MORE REAGENT

window is displayed on the screen. The screen states the reagent vial position, amount
needed and reagent that is missing. A 60 second count begins to count down and an
alarm sounds.

2. If the user notices the MORE REAGENT alarm and clicks the OK button, the ADD

REAGENT dialog box is displayed. This box contains a message notifying the user how
much reagent needs to be added.

3. The specified reagent volume is added and the user clicks the OK button.

4. The probe returns to the reagent vial and picks up the reagent. The Run Log records the

pause to add reagent.

Slides Skipped

Reagent vials are loaded on the Autostainer Plus and a run has been started.

1. The reagent probe senses insufficient or a missing reagent. The MORE REAGENT

window is displayed on the screen. The screen states the reagent vial position, amount
needed and reagent that is missing.

2. A 60 second count begins to count down and an alarm sounds.

3. If the user does not notice the MORE REAGENT alarm, after 60 seconds the RUN LOG

screen appears and the pause in the run is recorded.

4. The Autostainer Plus tries again to pump reagent into the probe. If no reagent is found,

this reagent is skipped and the Autostainer Plus continues with the next reagent. The Run
Log records no reagent was found and reports the skipped slides.

5. When the run is completed, the user is prompted with a list of slide(s) that have been

skipped. This list can be printed by selecting PRINT.

Partial Reagent Application

Reagent vials are loaded on the Autostainer Plus and a run has been started.

1. The reagent probe senses insufficient or a missing reagent. The MORE REAGENT

window is displayed on the screen. The screen states the reagent vial position, amount
needed and reagent that is missing.