Section 14 |

Starting a Run


After the specimen-containing slides and reagent vials have been loaded and scanned, it is advisable
to go through the following checklist:

Slides

• should be flooded with buffer containing 0.05% Tween 20 using a squirt bottle to prevent

slides from drying while loading additional slides.

• should be seated securely in slide rack.

Racks

• should be properly seated in the Autostainer Plus.

Reagents

• should be at room temperature.

• should be placed at proper locations in the reagent racks.

• should contain no air bubbles.

• should contain at least the reagent volume in each vial as specified in the Reagent Layout

Map.

Buffer

• reservoir should contain at least the volume of buffer indicated in the Reagent Layout Map

plus any additional volume listed for the completion of a delayed run.

Water

• reservoir should contain at least the volume of water indicated in the Reagent Layout Map.

Waste

• containers should have enough room to accommodate the volume of buffer and water used

during the staining run.


After the above checks have been made the Start Run sequence can be initiated.

The Dako Autostainer Plus enables the user to add Priority Slides during a run. See instructions later
in this section.

Starting a Run without Delay

Starting a staining run immediately.

1. When the bar code feature is disabled, select the NEXT button on the SLIDE LAYOUT

MAP:LOAD SLIDES screen during slide loading. The SET START TIME window is
displayed. When the bar code feature is enabled, the SET START TIME window is
displayed immediately after the slides have been scanned in the LOAD AND SCAN
SLIDES screen.