Section 3 |

Installation Requirements


For installation of the Dako Autostainer Plus, the following requirements should be met to ensure that
the instrument functions properly.

Requirements

Surface

Prior to unpacking the Dako Autostainer Plus, ensure that the
area to be used for instrument is a solid, level surface that can
safely support the weight of the system (approximately 240 lbs.
for instrument plus computer).

Size

Minimum dimensions for the Dako Autostainer Plus work area
are
2.00 m wide x 0.8 m deep x 1.0 m height (78” w x 30” d x 37” h)
clearance.

Environment

The work area must also be an environment with an ambient
temperature between 18° C-26° C (64° F-79° F), not facing
direct sunlight and not affected by a hood or vent.

Buffer Container

1 ten-liter plastic container.

Deionized Water Container

1 ten-liter plastic container (not available with the “01” hardware
version of the Dako Autostainer Plus).

Waste Containers

2 ten-liter plastic containers.

User Guide

1 User Guide with detailed operating instructions.

Seymour Glass Label
Printer

The Seymour slide labeling system is a simple, efficient label
printing system for histology, cytology and research laboratories.
This system allows you to print only the labels you need – when
you them – for samples, microscope slides, reagent vials and
other items.


Recommendations

Power source

Dako recommends a dedicated power source for the Dako
Autostainer Plus to prevent interference from other instruments
or equipment.