Assigning reagent drop zone, Cleaning schedule – Dako Autostainer Plus User Manual

Page 31

Advertising
background image

Dako Autostainer Plus

|

User Guide

31

Select the CANCEL button to escape without saving. The MAIN MENU screen is
displayed.

NOTE: The dispense volume selected on this screen is the default volume assigned to
all reagents in any programmed staining run. The volume for all reagents as well as for
specific reagents applied to all slides in a particular staining run can be changed in the
PROTOCOL TEMPLATE DESIGN screen. In addition, the dispense volume can also be
changed on an individual slide in the PROGRAM SLIDES screen (see Programming
Slides, Section 10).

4. To change the Default Reagent Dispense volume, click on the raised diamond next to the

new volume. A smaller black diamond appears next to the selected volume.

Assigning Reagent Drop Zone

The area(s) on the microscope slide where the reagent is dispensed is defined as the Drop
Zone(s). The default drop zone information is used when a new Protocol Template is created.

1. Move the mouse pointer over the top, middle or bottom portion of the microscope slide

image and press the left mouse button. A yellow bar appears on the slide in the selected
position. During the staining run, the selected Default Reagent Volume is dispensed on
this zone of the slides for all steps of the protocol.

2. Multiple drop zones can be selected by clicking on multiple (2 or 3) areas while holding

down the CONTROL (CTRL) key on the keyboard. Multiple zones are colored yellow.
During the staining run the selected Default Reagent Volume is dispensed on each of the
selected zones.

3. Select the OK button to save and the MAIN MENU screen is displayed.

Select the CANCEL button to escape without changing the Drop Zone. The MAIN MENU
screen is displayed.

NOTE: The drop zones(s) are selected to correspond to the location of tissues on the
slide.

Cleaning Schedule

This selection determines the number of slides the instrument will stain prior to prompting the
user to start a cleaning cycle. Dako recommends setting the cleaning schedule to 150 slides. The
MAIN MENU screen always displays the total number of slides processed since the last cleaning
cycle in red.

1. Move and click the mouse pointer on the Slides/Cleaning box and type in the number of

slides to stain between maintenance cleanings. Select the OK button and the MAIN
MENU screen is displayed.

Advertising