Reagent labels – Dako Autostainer Plus User Manual

Page 62

Advertising
background image

62

Dako Autostainer Plus

|

User Guide

1. From the PROGRAMMING GRID, select the PRINT button. The WHICH REPORT?

screen appears.

2. Select the SLIDE LABELS button. The PRINT SLIDE LABELS window appears.

3. To print all slide labels for the program, select the ALL LABELS button. All slide labels for

the current program will be printed. The format of each slide label is determined by the
SLIDE LABEL DESIGN screen that is accessed from the INITIALIZE screen (see
Initializing the Dako Autostainer Plus, section 6).

To print a range of labels, type the range of labels with a hyphen between the starting and
ending numbers in the range (i.e., 5-12). To print individual labels, type the label number
with commas between each number (i.e., 1,14,22).

These two formats can be used together to print a combination of a range of labels with
individual labels (i.e., 5-12,14,22).

NOTE: When entering individual or a range of labels, do not use a space between the
numbers.

NOTE: If you are adding a Priority Slide, select the UNPRINTED LABELS option in the
PRINT SLIDE LABELS screen that follows the programming grid when the NEXT button
is selected, and the system will print the Priority Slide label(s).

Reagent Labels

Printing reagent labels from a programmed run.

1. Program or retrieve a previously saved run on the PROGRAMMING GRID, select the

PRINT button. The WHICH REPORT? screen appears.

2. Select REAGENT LABELS button. The PRINT REAGENT LABELS screen appears

displaying all reagents from the programmed run on the PROGRAMMING GRID.

3. To print all reagent labels, select the PRINT ALL button. To print only selected reagents,

select desired reagents by positioning the mouse pointer over a reagent and clicking the
left mouse button. To select more than one reagent, repeat this step.

To de-select a selected reagent, position the mouse pointer over the highlighted reagent
and click with the left mouse button. The reagent will become un-highlighted, or de-
selected.

Once all desired reagents are highlighted, select the OK button.