Deleting an auto programming item, Using an auto programming item – Dako Autostainer Plus User Manual

Page 58

Advertising
background image

58

Dako Autostainer Plus

|

User Guide

row is high-lighted. Release the mouse button and the Save Auto Program window is
displayed.

NOTE: An Auto Program can consist of one or more rows of programming.

6. Enter a file name for the saved protocol. The Auto Program file name must be 8 alpha-

numeric characters or less; symbols and spaces cannot be used. Select the OK button.
The PROGRAMMING GRID appears.

7. Select Setup from the menu bar. Setup changes to Auto and exits the Auto program

function.

NOTE: To use a saved “Auto program,” the current template being programmed must
contain the same protocol elements listed in the same order as the saved Auto Program.
To find the correct protocol template, use the Auto file function. (See Using a Saved
Protocol Template for a specific Auto Program, earlier in this section). The dispense
volumes do not have to be the same.

Deleting an Auto Programming Item

1. Select Auto from the menu bar on the PROGRAMMING GRID.

2. Select Setup from the Auto menu. The word Auto changes to Setup.

3. Select any activated tile in the PROGRAMMING GRID to display the SAVE AUTO

PROGRAM window.

4. Use the mouse to scroll through the list of available Auto Programs. Select the Auto

Program to be deleted by clicking on it with the mouse.

5. Click on the DELETE PROGRAM button.

Select the YES button to delete the Auto Program and return to the SAVE AUTO
PROGRAM window.

Select the NO button to cancel changes and return to the SAVE AUTO PROGRAM
window.

Using an Auto Programming Item

1. Select Auto from the menu bar in the PROGRAMMING GRID.

2. Select Program from the Auto menu. The word Auto changes to Program.

3. Select a tile in the row where the Auto program is to be inserted. A list displays the saved

Auto Program files that can be assigned for this slide.