Delaying the start of a run, Starting a staining run at a later time – Dako Autostainer Plus User Manual

Page 85

Advertising
background image

Dako Autostainer Plus

|

User Guide

85

The current step, total and current time are always displayed in blue at the bottom of the
Run Log.

NOTE: If the current time is not accurate do not change it during the run. Adjust the time
at the completion of the run. To reset time refer to the Windows 98 or XP documentation.

NOTE: After the completion of the run, the slides will be rinsed with deionized water
every 60 minutes until the slides are removed from the Autostainer Plus (the deionized
water feature is not available for the “01” hardware version of the Dako Autostainer Plus).

Delaying the Start of a Run

Starting a staining run at a later time.

1. Select the gray scroll bar and hold the left mouse button down. Move the scroll bar to the

right until the desired Start Time and Finish Time is displayed. Release the mouse button.
The staining run will start and finish at the times indicated.

NOTE: The total amount of buffer is calculated and displayed below the scroll bar.

2. Select the PRIME PUMP (Water) button. A dialog box appears asking if the pathway is

clear for the arm to move.

Select the YES button. The Autostainer Plus calibrates the reagent dispensing
mechanism and primes the water-dispensing pump.

Select the NO button and the SET START TIME window is displayed.