Electronic checking and adjusting, continued, Checking and adjusting – Leica Geosystems TPS1100 Professional Series - User Manual User Manual

Page 31

Advertising
background image

31

TPS1100 - User Manual 2.2.1en

Checking and adjusting

6

10

12

21

Calibration Log-File

If the Log-File is enabled [LOGF+],
the measurements and results of the
calibration are stored in an ASCII-file.
This file is created in the directory
LOG on the memory card.
Subsequently, you can read the
memory card on your PC and obtain
a hard copy of the Log-file.

New data is always
appended to the calibration

Log-file.

TPS1100 - Instrument Calibration
Instrument

: TCRA1102plusSerial 619216

Compensator Index Error l,t
Date/Time

:

03/04/2000 15:43

Old values

:

l=

0.0000g

t= 0.0000g

Measurements

L= -0.0126g

T= 0.0298g

L= 0.0368g

T= 0.0164g

New values

:

l=

0.0010g

t= 0.0023g

Vertical Index Error i
Date/Time

:

03/04/2000 15:45

Old value

:

i=

0.0000g

Measurements

Hz= 377.0597 g

V= 104.2828 g

Hz= 177.0562 g

V= 295.7176 g

New value

:

i=

0.0001g

Example of a Calibration Log-file (here compensator and vertical index errors)