Electronic checking and adjusting, continued, Checking and adjusting – Leica Geosystems TPS1100 Professional Series - User Manual User Manual
Page 31

31
TPS1100 - User Manual 2.2.1en
Checking and adjusting
6
10
12
21
Calibration Log-File
If the Log-File is enabled [LOGF+],
the measurements and results of the
calibration are stored in an ASCII-file.
This file is created in the directory
LOG on the memory card.
Subsequently, you can read the
memory card on your PC and obtain
a hard copy of the Log-file.
New data is always
appended to the calibration
Log-file.
TPS1100 - Instrument Calibration
Instrument
: TCRA1102plusSerial 619216
Compensator Index Error l,t
Date/Time
:
03/04/2000 15:43
Old values
:
l=
0.0000g
t= 0.0000g
Measurements
L= -0.0126g
T= 0.0298g
L= 0.0368g
T= 0.0164g
New values
:
l=
0.0010g
t= 0.0023g
Vertical Index Error i
Date/Time
:
03/04/2000 15:45
Old value
:
i=
0.0000g
Measurements
Hz= 377.0597 g
V= 104.2828 g
Hz= 177.0562 g
V= 295.7176 g
New value
:
i=
0.0001g
Example of a Calibration Log-file (here compensator and vertical index errors)