39
TPS1100 - User Manual 2.2.1en
Checking and adjusting
6
10
12
21
Confirms readiness to
measure and the display
changes as follows.
Sight the target accurately.
Performs the second
measurement of the horizon-
tal angle.
After the second measurement has
been completed, the old and newly-
determined tilting-axis errors (a) are
displayed.
Stores the new values.
Repeats the complete tilting-
axis error determination
procedure.
Leaves the old values
unchanged.
If the value of the tilting axis
error (a) exceeds 5' 24"
(0.1 gon), the measurements are to
be repeated. If this happens often,
please notify service.
Tilting axis, continued
Main\ Tilting-Axis Error
old new
a Tilt: -0°00'03" 0°00'17"
Accept new value(s)?
YES RETRY
NO
MC
QUIT
Telescope Positioning
Hz- and V-positioning:
set direction(s) to zero.
∆∆∆∆∆
Hz
:
0°00'00"
∆∆∆∆∆
V
:
0°00'00"
OK
ABORT
MC
QUIT
Main\ Tilting-Axis Error
Aim accurately at the same
target in other face
Hz
:
193°19'24"
V
:
269°08'45"
MEAS
MC
QUIT