Checking and adjusting – Leica Geosystems TPS1100 Professional Series - User Manual User Manual

Page 40

Advertising
background image

40

TPS1100 - User Manual 2.2.1en

Checking and adjusting

6

0

2

21

By using the key

in the display

on page 30, it is possible to
determine the errors for vertical
index, line of sight and tilting axis (i/c/
k) with a single operation.

The V-index and line-of-sight errors
can be determined using a common
target which does not lie more than ±
9° (± 10 gon) away from the horizon-
tal plane. The determination of the
tilting-axis error requires a target
which lies at least ± 27° (± 30 gon)
above or beneath the horizontal
plane.

For exact procedural details,
please refer to the previous
chapters.

Combined error determination

Also the mechanical instrument-error
correction can be deactivated if the
need is to display and record only
raw data. To deactivate, set the
compensator correction and the
horizontal-circle correction to OFF,
i.e. the vertical angles relate to the
vertical axis and the horizontal
corrections are not taken into
account.

Deactivating the instrument-error correction

Main\

Compensator

Compensator ON/OFF

Hz corrections ON/OFF

compensat.:

ON

Hz-corr. :

ON

MC

QUIT

Advertising