Checking and adjusting

6

10

12

21

ATR collimation, continued

Stores the new values.

Repeats the ATR1 error
determination procedure.

Leaves the old values
unchanged.

No further repeat measure-
ments are required. The old

and the newly-defined ATR-collima-
tion errors become optional, and are
displayed together with the line-of-
sight errors (c) and the V-index errors
(i).

The calibration can be
repeated as often as

necessary until the required level of
accuracy is obtained. The result is the
mean of all the measurements taken.
It is recommended that at least two
measuring sequences be carried out.

The calibration process is
interrupted. The old values

will be left intact.

If the value of the ATR hori-
zontal and vertical

collimation errors exceeds 2' 42"
(0.05 gon), repeat the measurement
procedure.
Similarly, if the value of the V-index
error (i) exceeds 54' (1 gon) or if the
value of the line-of-sight error (c)
exceeds 5' 24" (0.1 gon) , the
measurements are to be repeated.
If these values are often exceeded,
please contact service.

CAL\ ATR Accuracy

No. of Meas:

2

σσσσσ

ATR Hz

:

0°00'05"

σσσσσ

ATR V

:

-0°00'08"

σ

σ

σ

σ

σ

i V-Ind. :

-----

σ

σ

σ

σ

σ

c Hz-col :

-----

Take more measurements?

CONT

MORE

ABORT

MC

CAL\ATR Collim. Errors

old new

ATR Hz

: 0°00'08" 0°00'05"

ATR V

: 0°00'10" 0°00'09"

i V-ind

: 0°00'00" 0°00'10"

c Hz-c

: 0°00'10" 0°00'02"

Accept new value(s)?

YES RETRY

NO

MC