Checking and adjusting – Leica Geosystems TPS1100 Professional Series - User Manual User Manual

Page 42

Advertising
background image

42

TPS1100 - User Manual 2.2.1en

Checking and adjusting

6

0

2

21

Toggles between simple
and combined error

determination.

YES =

Simultaneous
determination of ATR-
collimation error, line-of-
sight error and
vertical-index error

NO

=

Only determination of
ATR-collimation

It is advisable to determine
the ATR-collimation error,

the line-of-sight error and the vertical-
index error at the same time.

If the differences between the hori-
zontal and vertical angles exceed
±27' ( ±0.5 gon ), the display gives an
error message. The user is made
aware of this by an acoustic signal
and the

key is redefined as

"OK".
The measurement procedure can be
repeated.

When the second measurement has
been taken, the accuracy of ATR
and, if previously selected, the
accuracy of the index and the line-of-
sight errors are displayed.

ATR collimation, continued

Changes face automatically
directly after completing the initial
measurement.

Sight the prism accurately with the
crosshair.

Performs the second-face
measurement of the

collimation errors.

CAL\ ATR Collim. Errors

Aim accurately at the same

target in other face.

Hz

:

253°25'36"

V

:

271°14'46"

MEAS

MC

QUIT

Advertising