Checking and adjusting – Leica Geosystems TPS1100 Professional Series - User Manual User Manual

Page 33

Advertising
background image

33

TPS1100 - User Manual 2.2.1en

Checking and adjusting

6

10

12

21

After the first measurement of tilt, the
menu for non-motorized instruments
is the following:

Turn the instrument through 180°
(200 gon) so that

Hz = 0° 00' 00"

(0.0000 gon).

Compensator , continued

If the differences between the hori-
zontal and vertical angles lie within
±4° 30' (±5 gon), the display can be
quitted with

.

The user is made aware by an
acoustic signal that the

-key is

redefined as "OK".

Activates the second tilt
measurement.

Terminates the determination
of the compensator indexes.

The following dialog shows the two
newly-determined values for the
longitudinal and transverse
compensator-index errors.

Stores the new values.

Repeats the complete
calibration procedure.

Leaves the previous values
unchanged.

If the values for the index
errors (l, t) exceed 5' 24"

(0.1 gon), the complete calibration
procedure repeat, but first check that
the instrument is correctly levelled
and is free of vibration. If these
values are exceeded repeatedly,
please notify service.

Telescope Positioning

Hz- and V-positioning:

set direction(s) to zero.

Hz

:

0°00'00"

V

:

------

OK

ABORT

MC

QUIT

II

Telescope Positioning

Hz- and V-positioning:

set direction(s) to zero.

Hz

:

180°00'00"

V

:

------

OK

ABORT

MC

Main\ Compens.Index Error

new

l Comp:

0°03'02"

t Comp:

-0°02'06"

Accept new value(s)?

YES RETRY

NO

MC

QUIT