English – Compex Energy mi-Ready User Manual

Page 166

Advertising
background image

164

English

T

he duration of the charging operation is

shown on the screen (a complete charge
may take 10 hours).

T

he battery icon is animated while the

battery is recharging. When fully charged,
the icon will be full and the total time
taken to recharge the battery will flash on
the screen. Simply disconnect the charger:
the Compex will turn off automatically.

9. Problems and solutions

Electrode fault

T

he Compex "bleeps" and alternatively

displays the symbol of a pair of
electrodes and an arrow pointing to the
channel where a problem has been
detected. In the example above, the
stimulator has detected an error in
channel 1.

Check that electrodes are connected
to this channel.

Check whether the electrodes are old,
worn, and/or the contact is poor: try
using new electrodes.

Try using the stimulation cable on a
different channel. If the cable is still
showing a fault, replace it.

Stimulation is not producing
the usual sensation

Check that all the settings are correct
and ensure the electrodes are
positioned properly.

Change the positioning of the
electrodes slightly.

The stimulation effect causes
discomfort

The electrodes are beginning to lose
adhesion and no longer provide good
contact on the skin.

The electrodes are worn and need to
be replaced.

Change the positioning of the
electrodes slightly.

The stimulator is not working

If an error screen appears while you
are using the device, note the error
number (in the example in fig. 12,
error 1/0/0) and contact the nearest
authorised customer support service.

ENERGY2 FR DE IT ES EN:ENGY Man FR 31/01/08 14:36 Page 164