Compex Energy is designed for use in the electromagnetic environment

stipulated below.

The buyer or user of the Compex Energy must ensure it is used

in this recommended environment.

RECOMMENDATIONS AND DECLARATION BY THE MANUFACTURER -

NOTE: VT is the AC supply voltage before application of the test
level.

Immunity test

Test level IEC 60601

Electrostatic discharge

(DES)

CEI 61000-4-2

± 6 kV at the contact

± 8 kV in air

Rapid electrical transients in

bursts of noise

CEI 61000-4-4

± 2 kV for power supply lines
± 1 kV for input/output lines

Surge currents

CEI 61000-4-5

± 1 kV differential mode

± 2 kV common mode

Voltage trough, short power cuts

and voltage variations on power

supply input lines

CEI 61000-4-11

< 5% VT (trough > 95% of UT)

for 0.5 cycle

< 40% VT (trough > 60% of VT)

for 5 cycles

< 70% VT (trough > 30% of VT)

for 25 cycles

< 5% VT (trough > 95% of VT)

for 5 seconds

Observance level

± 6 kV at the contact

± 8 kV in air

Not applicable

System battery-powered

Not applicable

System battery-powered

Not applicable

System battery-powered

Magnetic field

at mains frequency

(50/60 Hz)

CEI 61000-4-8

3 A/m