English

0,01

0,12

0,1

0,38

1

1,2

In the case of transmitters whose maximum output power is not shown in the table above, the recommended

spacing of d metres (m) can be calculated using the appropriate equation for the transmitter frequency, where

P is the maximum output power of the transmitter in watts (W) as set by the transmitter manufacturer.

The Compex Energy is designed for use in an electromagnetic environment in which

radiated RF waves are controlled. The buyer or user of the Compex Energy can

contribute to preventing electromagnetic interference by maintaining a minimum

distance between RF portable and mobile communication appliances (transmitters) and

the et Compex Energy according to the table of recommendations below and according

to the maximum output power of the telecommunication appliance.

RECOMMENDED SPACING BETWEEN A PORTABLE AND MOBILE

COMMUNICATION APPLIANCE AND THE COMPEX ENERGY

NOTE 1: At 80 MHz and at 800 MHz, the spacing

for high frequency amplitude is applied.

NOTE 2: These guidelines may not be appropriate for some

situations. Electromagnetic wave propagation is modified by

absorption and reflection due to buildings, objects and persons.

0,12

0,23

0,38

0,73

1,2

2,3

10

3,8

100

12

3,8

7,3

12

23

Maximum transmitter

output power W

From 150 kHz

to 80 MHz

d = 1.2 P

From 80 MHz

to 800 MHz

d = 1.2 P

From 800 MHz

to 2.5 GHz

d = 2.3 P

Spacing according to the frequency of the transmitter mCISPR 11