Compex Energy is designed for use in the electromagnetic environment stipulated

below. The buyer or user of the Compex Energy must ensure it is used in this

recommended environment.

RECOMMENDATIONS AND DECLARATION BY THE MANUFACTURER -

NOTE: VT is the AC supply voltage before application of the test
level.

Immunity test

Test level IEC 60601

RF immunity

IEC 61000-4-6

3 Vrms

150 kHz to 80 MHz

Radiated RF

IEC 61000-4-3

3 V/m

80 MHz to 2.5 GHz

10 V/m

26 MHz to 1 GHz

Observance level

Signal lines

Not applicable for power supplies

of battery-powered systems

3 A/m

10 V/m

NOTE 1: From 80 MHz to 800 MHz, the spacing for high frequency amplitude is applied.

NOTE 2: These guidelines may not be appropriate for some situations.

a

The field intensity from fixed transmitters, such as radio telephone base stations (cellular/wireless) and a

mobile radio, amateur radios, AM and FM radio emissions and TV emissions cannot be predicted with any

accuracy. It may therefore be necessary to consider an analysis of the electromagnetic environment of the

site to calculate the electromagnetic environment coming from RF fixed transmitters. If the field intensity

measured in the environment where the Compex Energy is located exceeds the appropriate RF observance

level above, the Compex Energy should be monitored to ensure it is operating properly.

ENERGY2 FR DE IT ES EN:ENGY Man FR 31/01/08 14:36 Page 180