Limit – KEYENCE LS-3100/3100W Series User Manual

Page 29

Advertising
background image

23

Example of Operation

You are to measure the diameter of round bars. The reference diameter is 10
mm, and you wish to convert a difference of 1 mm into 1 V.
In this case, first set 100.000 mm to 0 V. Then, to represent each 1 mm
change in diameter from the reference by 1 V, set the reference plus 6 mm to
+6 V and the reference minus 6 mm to -6 V. That is, +6 V = 16 mm and -6 V
= 4 mm.

Step

Key operation

Display

1

Use PRM to display analog
voltage range.

2

Press "1".

3

Press "6".

4

Press ".".

5

Press "0" four times.

6

Press ENT

7

Press "4".

8

Press ".".

9

Press "0" four times.

10

Press ENT.

• The LS controller is factory-set to +6 V = 60 mm and -6 V = - 60 mm.
• A range of 0.024 mm to 240 mm can be specified.
• Make sure that the value corresponding +6 V is greater than that for -6 V.
• When your reference value is an integer with no decimals, steps 4, 5, 8

and 9 can be omitted.

CHAPTER 4 Functions and Controls

+ 6

6 0 . 0

V

0 0 0

-

=

-

0 . 0 0 0 0

=

V

6

6

+ 6 V

1

-

=

6

. 0 0 0 0

=

V

6

0

[

]

-

+ 6 V

6

-

=

6

. 0 0 0 0

=

V

6

0

[

]

-

1

+ 6 V

-

=

6

. 0 0 0 0

=

V

6

0

[

]

-

1

. 0 0 0 0

6

+ 6 V

-

=

=

V

6

1

. 0 0 0 0

6

+ 6 V

-

=

4

=

V

6

1

. 0 0 0 0

6

[

]

+ 6 V

-

=

.

=

V

6

1

. 0 0 0 0

6

[

]

4

+ 6 V

-

=

=

V

6

1

. 0 0 0 0

6

[

]

4 . 0 0 0 0

+ 6 V

-

=

=

V

6

1

. 0 0 0 0

6
4 . 0 0 0 0

+ 6 V

6

-

=

6

. 0 0 0 0

=

V

6

0

[

]

-

1

.

LIMIT

You can set the following 3 parameters in LIMIT:

• 3-step or 7-step differentiation
• Tolerance limits when 7-step differentiation is selected
• Comparator output type; Relay or open-collector (if the measured value is

above the upper limit or below the lower limit).

• Use FNC to set the tolerance limits for the measured value appearing on

the subdisplay (CH2).

➪ See P.32