KEYENCE LS-3100/3100W Series User Manual

Page 87

Advertising
background image

81

Check

item

Problem

Details of inspection

Corrective action

Reference

page

CHAPTER 9 Troubleshooting Guide

Abnormal measure-
ment values are being
displayed.

• Do the serial numbers match

between the controller and
scanning head?

• Has calibration been

correctly performed?

• Is the laser beam correctly

aligned?

• Is the target tilted or has it

moved from the original
position?

• Are the targets correctly

placed within the measure-
ment area?

• Is a reference edge placed

within the measurement
area?

• Is there any dust or dirt on

the target?

• Isn’t the controller set to the

analog output range setting
mode?

• Connect a scanning head

with the matching serial
number.

• Correctly perform calibration.

• Check whether or not the

laser beam is correctly
aligned in the laser beam
alignment check mode. If the
beam is not correctly aligned,
refer to the error message as
indicated in the display and
adjust the mounting condi-
tions accordingly.

• Correctly place the targets.

• Correctly place the targets

within the measurement area.

• Place a reference edge within

the measurement area.

• Remove any dust or dirt from

the target.

• Set the appropriate measure-

ment mode.

p. 14

p. 41

p. 65

"H-Err" is displayed.

Is the scanning head
properly connected?

Turn the power OFF, then after
correctly connecting the
scanning head, turn the power
ON.

Analog
output

No output.
(0 V output)

Is the measurement value
displayed?

To make sure that the target is
correctly measured, correctly
position the controller and
targets.

Voltage above +6 V or
below -6 V is output.

Is the analog output range
correctly set?

Check that the analog output
range is correctly set.

p. 78

• Are the targets correctly

placed in the measurement
area?

• Is a standard edge placed

within the measurement
area?

• Is there any dust or dirt on

the laser aperture?

• Is there any dust or dirt on

the targets?

• Is there a large fluctuation of

ambient temperature?

• Correctly place the target

within the measurement area.

• Place a standard edge within

the measurement area.

• Remove any dust or dirt.

• Remove any dust or dirt.

• Maintain the constant operat-

ing temperature.

The measurement display
value contains unusual
dispersion.

Measure-
ment
Value
Display

p. 22

p. 84

08.10.1, 10:57 AM

81