2 gp-ib interface i/o, Specifications, Pin assignment – KEYENCE LS-3100/3100W Series User Manual

Page 76: 2. gp-ib interface i/o

Advertising
background image

70

When EXT TRIG IN is pulled low with ETE IN pulled low, the measured value
is output after a max. of 20 ms (MOVE averaging, number of measurements
for averaging: 16 to 1,024), or a max. of "2.5 ms x number of measurements
for averaging" (SIMPLE averaging). The output value is retained until
EXT TRIG IN is pulled low again.

External Trigger Input
(EXT TRIG IN)

6-2. GP-IB Interface I/O

Specifications

The GP-IB I/O interface conforms to IEEE (Institute of Electrical and Elec-
tronics Engineers) standard 488-1978.

Transmission code

Logic level

Number of connectable devices
Total cable length

Cable length between 2 devices

ASCII

15 max. (including controller)

20 m max.*

4 m max.

*

Varies according to the number of interconnected devices.

Pin Assignment

Pin No.

Signal

Function

1

DIO1

Data bus

2

DIO2

Data bus

3

DIO3

Data bus

4

DIO4

Data bus

5

EOI

Indicates that the data transmission is
completed.

6

DAV

Indicates that the data is valid.

7

NRFD

Indicates that the data reception is impossible.

8

NDAC

Indicates that the data reception is not
completed.

9

IFC

Initializes the interface.

10

SRQ

Allows devices on the line to request interrupt to
the controller.

11

ATN

Determines whether the data is interface
message or device message.

12

Shield

Shield

13

DIO5

Data bus

14

DIO6

Data bus

15

DIO7

Data bus

16

DIO8

Data bus

CHAPTER 6 Optional Interface Boards

Logic 0 (HIGH state): +2.4 V min.
Logic 1 (LOW state): +0.4 V max.

(Bidirectional transceiver is used.)