Disp (subdisplay) key, Ofs (offset) key, Disp (subdisplay) key ofs (offset) key – KEYENCE LS-3100/3100W Series User Manual

Page 49: Ee disp (subdisplay) key, Ff ofs (offset) key

Advertising
background image

43

CHAPTER 4 Functions and Controls

4

Enter "20.1000".

5

Press ENT to register.
"CALIB OK" will appear.

6

Press CAL to display
"CAL Y".

7

Repeat steps 2 to 5 to
calibrate Y.

E

E

E

E

E DISP

(Subdisplay) Key

One of the following items appears on the subdisplay:
• Status of settings during panel key operation

• Measured value for CH1 in NORMAL, P-P, PEAK, or BOTTOM mode

Measured value in any of the desired mode can be displayed regardless
of the mode set using MODE. The comparator output and BCD output will
be sent based on the mode set using MODE.

• Current settings for CH1

Current parameters or modes set using SEG, MODE, or AVE can be
displayed.

• Measured values for CH2

Each time you press DISP, the values or parameters mentioned in (2), (3),
and (4) will display alternately.

* These values or parameters cannot be changed on the subdisplay.

F

F

F

F

F OFS (Offset) Key

Offset value is set in this mode. This value is added to the measured value,
and the total value is displayed and output. Any value between -999.9999
and 9999.9999 can be specified as an offset value. If the diameter of a round
bar or width of a sheet material is too large to measure with one scanning
head, offset value can be used to measure it using two scanning heads. In
this case, the diameter or width is displayed or output in absolute value.

* Use FNC to set the offset value for CH2.

➪ See P.39.

C

L

0

X

0 1 0

.

A

0

0

1 0 0

.

0

1

[ 2

]

C

L

B

A

K

I

O

C

L

0

Y

0 0 0

.

A

0

0

0 0 0

.

0

Procedure

1. Press OFS.

2. Enter the desired offset value.

3. Press ENT to register.
Example of Operation

08.10.1, 10:55 AM