Temperature characteristics (typical) – KEYENCE LS-3100/3100W Series User Manual

Page 91

Advertising
background image

85

Temperature
Characteristics
(Typical)

When a rod 10 mm in diameter is measured at the center of the measuring
area [reference temperature: 20°C (68°F)].

LS-3032

Temperature

0°C (32°F)

10°C (50°F) 20°C (68°F) 30°C (86°F) 40°C (104°F)

Drift

+2.2 µm

+1.1 µm

0

-1.4 µm

-3.2 µm

LS-3033*

Temperature

0°C (32°F)

10°C (50°F) 20°C (68°F) 30°C (86°F) 40°C (104°F)

Drift

+2.0 µm

+0.8 µm

0

-1.0 µm

-2.2 µm

LS-3034

Temperature

0°C (32°F)

10°C (50°F) 20°C (68°F) 30°C (86°F) 40°C (104°F)

Drift

+2.6 µm

+1.3 µm

0

-1.4 µm

-3.2 µm

LS-3036

Temperature

0°C (32°F)

10°C (50°F) 20°C (68°F) 30°C (86°F) 40°C (104°F)

Drift

+3.7 µm

+1.7 µm

0

-1.7 µm

-3.5 µm

When a rod 20 mm in diameter is measured at the center of the measuring
area [reference temperature: 20°C (68°F)].

LS-3060

Temperature

0°C (32°F)

10°C (50°F) 20°C (68°F) 30°C (86°F) 40°C (104°F)

Drift

+8.6 µm

+4.3 µm

0

-4.2 µm

-8.4 µm

* Contact us for the data of LS-3033SO (8073)

APPENDIX B Characteristics