Temperature characteristics (typical) – KEYENCE LS-3100/3100W Series User Manual
Page 91
85
Temperature
Characteristics
(Typical)
When a rod 10 mm in diameter is measured at the center of the measuring
area [reference temperature: 20°C (68°F)].
LS-3032
Temperature
0°C (32°F)
10°C (50°F) 20°C (68°F) 30°C (86°F) 40°C (104°F)
Drift
+2.2 µm
+1.1 µm
0
-1.4 µm
-3.2 µm
LS-3033*
Temperature
0°C (32°F)
10°C (50°F) 20°C (68°F) 30°C (86°F) 40°C (104°F)
Drift
+2.0 µm
+0.8 µm
0
-1.0 µm
-2.2 µm
LS-3034
Temperature
0°C (32°F)
10°C (50°F) 20°C (68°F) 30°C (86°F) 40°C (104°F)
Drift
+2.6 µm
+1.3 µm
0
-1.4 µm
-3.2 µm
LS-3036
Temperature
0°C (32°F)
10°C (50°F) 20°C (68°F) 30°C (86°F) 40°C (104°F)
Drift
+3.7 µm
+1.7 µm
0
-1.7 µm
-3.5 µm
When a rod 20 mm in diameter is measured at the center of the measuring
area [reference temperature: 20°C (68°F)].
LS-3060
Temperature
0°C (32°F)
10°C (50°F) 20°C (68°F) 30°C (86°F) 40°C (104°F)
Drift
+8.6 µm
+4.3 µm
0
-4.2 µm
-8.4 µm
* Contact us for the data of LS-3033SO (8073)