CHAPTER 4 Functions and Controls

Table of I/O
Terminals

Analog Voltage Output

Analog voltage signal for CH1 can be obtained through the MON terminals.
The LS controller has been factory-set to convert a range from -60 mm to
+60 mm into a voltage range from -6 V to +6V. (No voltage signal is provided
for CH2. If voltage signal for CH2 is required, consult the KEYENCE distribu-
tor.)

Analog voltage range: A range between -6 V and +6 V can be set to
match a desired measurement range. Any value between -999.9999 and
9999.9999 can be selected. Note, however, that the difference between the
maximum and minimum measurements (those that you specify to corre-
spond respectively to +6 V and -6 V) must fall within a range between 0.024
mm and 240 mm.

Calibration: When the segment selectors are calibrated, the analog voltage
signal through these terminals and RS-232C interface is automatically sent
based on the calibrated value.

Offset: When the segment selectors are calibrated, the selected offset
value is added to the analog voltage signal sent through these terminals and
RS-232C interface.

The analog voltage which appears on the main display and that which is
output through the MON terminals are always the same.

Comparator Output

For CH1 (Channel 1)

HI3

HI2

HI1

GO

LO1
LO2

LO3

HIGH

GO

LOW

HIGH
GO

LOW

HI1
GO

LO1

HIGH
GO

LOW

HIGH
GO

LOW

HI

GO

LO

CH1

CH1

CH2

For CH2 (Channel 2)

CH2

7-step differentiation

3-step differentiation

Only 3-step
differentiation is
available.

Tolerance Limit

Output Device/Terminal

Control I/O (open-collectors)

Comparator output terminals (relays)

Front panel (LEDs)

Control I/O (open-collectors)

Comparator output terminals (relays)

Front panel (LEDs)

CH1

Control I/O (open-collectors)

08.10.1, 10:55 AM

47