5-1. Alignment of Optical Axis (LS-3060 and LS-3033 only)

The LS-3060 is delivered with the transmitter and receiver (scanning head)
mounted to the detachable frame. When using the LS-3060 with the transmit-
ter and receiver removed from the frame on when using the LS-3033, be
sure to align them in terms of the optical axis of the laser. (See the figure
below.) Misalignment of the optical axis will cause measurement errors.
(A special optical axis checker is available with the LS-3060 and LS-3033.
Refer to "5-2. Checking Alignment of Optical Axis" on page 66 for the align-
ment procedures.)

Allowance for
Aligning Transmitter
and Receiver

Deviation in height or width

Inclination in height or width

Within ±1 mm

Within ±1 mm

Within ±1 mm

Within ±1 mm

Transmitter

Receiver

Transmitter

Transmitter

Receiver

Transmitter

Receiver

Within ±0.25°

Within ±0.25°

Within ±0.25°

Within ±0.25°