Factory

Sequences

Introduction

B

B-7

Table B-2. Factory Sequences (MPS450 and Marq III) (cont.)

Dual
Keys Sequence

Segments/Selections

Run

Times

Serial

Command

Set ECG type to adult

0:00:00

ADULT

BP1 /Resp Artifact OFF

0:00:00

P1AOFF

BP2/Resp Artifact OFF

0:00:00

P2AOFF

BP3 /Resp Artifact OFF

0:00:00

P3AOFF

BP4/Resp Artifact OFF

0:00:00

P4AOFF

ECG/NSR BPM 80

0:00:00

NSR80

Respiration Rate 20

0:00:00

RR20

Zero ALL BPs

0:00:00

ZALL

Q

L

L

Monitor

Setup

STOP

Set ECG type to adult

0:00:00

ADULT

ECG/NSR BPM 80

0:00:00

NSR80

Respiration Ohms 1.0

0:00:00

RO1.0

Respiration Rate 20

0:00:00

RR20

BP1 ATM (0)

0:00:00

P1S0

BP2 ATM (0)

0:00:00

P2S0

BP3 ATM (0)

0:00:00

P3S0

BP4 ATM (0)

MANUAL

• P4S0

ECG/NSR BPM 240

0:00:00

NSR240

Respiration Rate 120

0:00:00

RR120

Q
L
M

Monitor

Test

STOP

Set ECG type to adult

0:00:00

ADULT

Set ECG type to adult

0:00:00

ADULT

Set ECG type to adult

0:00:00

ADULT

BP1 /Resp Artifact ON

0:00:00

P1AON

BP2/Resp Artifact ON

0:00:00

P2AON

BP3 /Resp Artifact ON

0:00:00

P3AON

Q
L
N

Monitor

Start

BP4/Resp Artifact ON

0:00:00

P4AON