Dual-key commands introduction, A-19, Table a-1. dual-key commands (cont.) – Fluke Biomedical HHC3 User Manual

Page 87

Advertising
background image

Dual-Key

Commands

Introduction

A

A-19

Table A-1. Dual-Key Commands (cont.)

Keys

Command Description

medSim 300B

MPS450

Marq III

Cardiac output Baseline temperature

36 deg C

COB36 (Step)

37 deg C

COB37

38 deg C

COB38

F
O

User adjusted

COBUSR

Cardiac output Waveform

Reset

CRSET (Step)

3 L/M flow rate

COW3

5 L/M flow rate

COW5

7 L/M flow rate

COW7

Interrupted injectate

COWINT

Left ventricular shunt

COWLVS

F
P





Slow injectate

COWSLW

Cardiac output Trend


Start sequence thru selected trend

COTSTRT
(Step)

Normal sequence

COTNRM

Sequence with interrupted injectate

COTINT

G
B

Sequence with slow injectate

COTSLO

Cardiac Output Injectate / Wave Form

Turn CO wave off.

CRSET CRSET

Run CO wave 0 degrees C at (2.5l/min) 2.7 l/min.

COI0, COW2.5
(Step)

COI0, COW2.5
(Step)

Run CO wave 0 degrees C at 5.0 l/min.

COI0, COW5.0

COI0, COW5.0

Run CO wave 0 degrees C (10.0l/min) 9.9 l/min.

COI0, COW10.0

COI0, COW10.0

Run CO wave 24 degrees C at 2.5 l/min.

COI24, COW2.5

COI24, COW2.5

Run CO wave 24 degrees C at 5.0 l/min.

COI24, COW5.0

COI24, COW5.0

Run CO wave 24 degrees C at (10.0l/min) 10.4
l/min.

COI24, COW10.0

COI24, COW10.0

Run CO wave faulty injectate 0 degrees C.

COI0, COWFLT

COI0, COWFLT

Run CO wave faulty injectate 24 degrees C.

COI24, COWFLT

COI24, COWFLT

Run CO wave left/right shunt 0 degrees C.

COI0, COWLRS

COI0, COWLRS

Run CO wave left/right shunt 24 degrees C.

COI24, COWLRS

COI24, COWLRS

Run CO wave cal pulse 0 degrees C.

COI0, COWCAL

COI0, COWCAL

J
P









Run CO wave cal pulse 24 degrees C.

COI24, COWCAL

COI24, COWCAL