Compliance information, Scanner instruction manual, Safety – Precision Digital PD6080 User Manual

Page 15: Electromagnetic compatibility

Advertising
background image

PD6080/PD6081 Super Snooper Modbus

Scanner Instruction Manual

15

COMPLIANCE INFORMATION

Safety

UL & c-UL LISTED

USA & Canada
UL 508 Industrial Control Equipment

UL FILE NUMBER

E160849

FRONT PANEL

UL Type 4X, NEMA 4X, IP65; panel gasket provided

LOW VOLTAGE
DIRECTIVE

EN 61010-1:2001
Safety requirements for measurement, control, and laboratory use

Electromagnetic Compatibility

EMISSIONS

EN 55022:2006/A1:2007
Class A ITE emissions requirements

Radiated
Emissions

Class A

AC Mains
Conducted
Emissions

Class A

IMMUNITY

EN 61326-1:2006
Measurement, control, and laboratory equipment
EN 61000-6-2:2005
EMC heavy industrial generic immunity standard

RFI - Amplitude
Modulated

80 -1000 MHz 10 V/m 80% AM (1 kHz)
1.4 - 2.0 GHz 3 V/m 80% AM (1 kHz)
2.0 - 2.7 GHz 1 V/m 80% AM (1 kHz)

Electrical Fast
Transients

±2kV AC mains, ±1kV other

Electrostatic
Discharge

±4kV contact, ±8kV air

RFI - Conducted

10V, 0.15-80 MHz, 1kHz 80% AM

AC Surge

±2kV Common, ±1kV Differential

Surge 1KV

(CM)

Power-Frequency
Magnetic Field

3 A/m 70%V for 0.5 period

Voltage Dips

40%V for 5 & 50 periods
70%V for 25 periods

Voltage
Interruptions

<5%V for 250 periods

Note:

Testing was conducted on PD6000 Series meters installed through the covers of grounded metal
enclosures with cable shields grounded at the point of entry representing installations designed to
optimize EMC performance.

Declaration of Conformity available at www.predig.com