Scanner instruction manual, Scanner copy function ( copy ) – Precision Digital PD6080 User Manual

Page 95

Advertising
background image

PD6080/PD6081 Super Snooper Modbus

Scanner Instruction Manual

95

Error Message (Error)

An error message indicates that the calibration or scaling process was not successful. After the error
message is displayed, the scanner reverts to input 2 during calibration or scaling and to input 1 during
internal calibration, allowing the appropriate input signal to be applied or programmed.

The error message might be caused by any of the following conditions:
1. Input signal is not connected to the proper terminals, or it is connected backwards.
2. Minimum input span requirements not maintained.

Input Range

Input 1 & Input 2 Span

4-20 mA

0.15 mA

±10 VDC

0.10 VDC


Scanner Copy Function (Copy)

The Copy function is used to copy (or clone) all the settings from one scanner to other scanners requiring
exactly the same setup and programming (i.e. Enabled PVs, display setup, scaling, decimal point, etc.).

Warning!

Only the PDA1200 meter copy cable must be used for scanner-to-
scanner copy function. Using standard CAT5 or other cable will cause
damage to both scanners.

Figure 20. Scanner Copy Connection


For information on Precision Digital cables and adapters, see page 8 for details.

Copy Function Requirements

To successfully copy settings from one scanner to another, both scanners must be set to Slave
Mode. Additionally, they must be the same model, have the same software version, and the
same baud rate setting.
See Determining Software Version, page 98, for instructions.

PDA1200 Meter Copy Cable

Advertising
This manual is related to the following products: