Troubleshooting, Scanner instruction manual, Troubleshooting tips – Precision Digital PD6080 User Manual

Page 97

Advertising
background image

PD6080/PD6081 Super Snooper Modbus

Scanner Instruction Manual

97

TROUBLESHOOTING

The rugged design and the user-friendly interface of the scanner should make it unusual for the installer
or operator to refer to this section of the manual. However, due to the many features and functions of the
scanner, it’s possible that the setup of the scanner does not agree with what an operator expects to see.
If the scanner is not working as expected, refer to the recommendations below.

Troubleshooting Tips

Symptom

Check/Action

No display at all

Check power at power connector

Not able to change setup or programming, Locd is displayed

Scanner is password-protected, enter correct six-digit password to unlock

Scanner displays error message during scaling (Error)

Check:

Input 2 must greater than Input 1, Input 3 must be greater
than Input 2, etc.

Scanner displays
1.

999999

2.

-99999

Indicating overrange or underrange condition
Check the input data value and scaling in Setup menu

Display stop scanning,

LED indicator flashing

Check:

1.

Stop on alarm feature has been enabled

2.

Press PREV or NEXT to resume scanning

Displays break message

Check:

1.

RS-485 connection to slave devices

2.

Slave Id, register number of slave devices

3.

Baud rate and parity of all devices on the bus

4.

Scanner Id must be different from other devices

Snooper mode not reading the PVs on the RS-485 bus

Check:

1.

Increase Master’s Transmit Delay
(e.g. Snooper delay = 100ms, Master delay = 110ms)

2.

Increase Snooper’s byte-to-byte timeout

3.

Decrease the slave device’s transmit delay to <10ms

4.

Snooper cannot read the same PV twice, check setup

Scanner experiencing faults and communication breaks

Check:

1.

Increase response time (t-resp) and/or transmit delay
(tr dly). This may require some trial and error, as these
are dependent upon the number of devices on the bus.

2.

Internal Scan ID Modbus address. Addresses 256 (mA)
or 257 (V) are used for Channel A, while Addresses 258
(mA) or 259 (V) are used for Channel B.

Scanner not communicating with ScanView software

Check:

1.

Serial adapter and cable

2. Serial

settings

3.

Scanner address, baud rate, and transmit delay

Display does not respond to input data, reading a fixed
number

Check:

Display assignment, it might be displaying max/ min

Display reading is not accurate

Check:

1. PV

Scaling

2.

Check format selected: Dec or Ft&In

Relay operation is reversed

Check:

1. Fail-safe

in

Setup

menu

2.

Wiring of relay contacts

Relay and status LED do not
respond to signal

Check:

1.

Relay action in Setup menu

2.

Set and reset points

Flashing relay status LEDs

Relays in manual control mode or relay interlock switches opened.

If the display locks up or the scanner does not respond at all

Cycle the power to reboot the microprocessor.

Other symptoms not described above

Call Technical Support for assistance.

Advertising
This manual is related to the following products: