Scanner instruction manual, Input calibration menu ( ical ) – Precision Digital PD6080 User Manual

Page 94

Advertising
background image

PD6080/PD6081 Super Snooper Modbus

Scanner Instruction Manual

94

Input Calibration Menu (ICAL)

To Run Mode if calibration is

successful, otherwise an

message will appear.

Recalibrating the Analog Input Channels (Ch-A & Ch-B)

The analog input channels are calibrated at the factory. If recalibration is needed, follow the flowchart
menu above and the instructions below.

1. Enter the ICAL menu to calibrate channel A and channel B.
2. Select mA or Volt
3. Apply the low (4 mA or 0 V) and high (20 mA or 10 V) signals as requested.
4. Press Enter to accept the values entered or press Menu to exit calibration without saving the

changes.

Advertising
This manual is related to the following products: